View
View
View
View
| Display | LCD, maximum 2000 counts, display refresh rate approximately 2 times/second |
| Type of Product | Earth Resistance Meter |
| Model No | FT6380-90 |
| Power Supply | AA alkaline batteries (LR6) x 2 |
Display |
LCD, maximum 2000 counts, display refresh rate approximately 2 times/second |
Type of Product |
Earth Resistance Meter |
Model No |
FT6380-90 |
Power Supply |
AA alkaline batteries (LR6) x 2 |
Country of Origin |
Japan |
Monotaro ID |
51213008 |
Alarm |
[Function] Resistance/Current Alarm: Beeps when below or above the threshold. |
Included Components |
Includes: Carrying case x 1, Test resistors (1Ω, 25Ω) x 1 each, Strap x 1, AA alkaline batteries (LR6) x 2, Instruction manual x 1, Precautions for use x 1 |
Name of Manufacturer/Packer/Importer |
HIOKI E.E. CORPORATION |
Current Range |
20mA (resolution 0.01mA) to 60A (resolution 0.1A), 5 ranges, zero suppression below 0.05mA. |
Import Country |
Direct Import from Japan |
Mass (g) |
Approximately 620 (excluding batteries) |
Specifications |
Maximum allowable input (current measurement): AC 100 A continuous, AC 200 A for 2 minutes (50 Hz/60 Hz), Harmonics: [When Z3210 is installed] Harmonic levels/content/total harmonic distortion up to the 30th order in current (displayable with application software (GENNECT Cross)) |
Dimension (WxDxH) |
Approximately 73x218x43 mm |
Dust/Water Protection |
IP40 (EN60529) when the clamp sensor is closed |
Maximum Rated Voltage |
[To Ground] AC 600 V Measurement Category IV (Expected transient overvoltage 8000 V) |
End To Ground Resistance |
0.2 Ω (resolution 0.01) to 1600 Ω (resolution 20), 10 ranges, zero suppression below 0.02. |
Basic Accuracy |
[Grounding resistance range] ±15%rdg ±0.02Ω, [Current range] ±20%rdg ±0.05mA (30Hz to 400Hz, true RMS rectification) |
Continuous Operating Time (hrs) |
Approximately 40 (measured at 25 Ω, backlight off, Z3210 not installed) / Approximately 35 (measured at 25 Ω, backlight off, Z3210 installed, wireless communication in progress) |
Measurable conductor diameter (Φmm) |
32 |
Operation Type |
Calculate the grounding resistance value by performing voltage injection and current measurement with an integrated clamp (dedicated to multiple grounding where one grounding line is grounded multiple times, the more grounding electrodes, the more accurate the value +) |
Crest Factor |
50 or less (60 A range is 17 or less) |
Other functions |
Data hold, backlight, filter, auto power save, wireless communication (only when Z3210 is installed) |
Memory Function |
2000 memory storage |