View
View
View
View
| Type of Product | Ac/Dc Current Sensor |
| Model No | CT6873-01 |
| Power Supply | Supplied via power analyzer PW8001, PW6001, PW3390, sensor unit CT9555, CT9556, CT9557, 3CH current unit U8977 |
| Output Voltage | Rate: 10mV/A *The output of this device is AC+DC voltage output via the sensor unit. |
Type of Product |
Ac/Dc Current Sensor |
Model No |
CT6873-01 |
Power Supply |
Supplied via power analyzer PW8001, PW6001, PW3390, sensor unit CT9555, CT9556, CT9557, 3CH current unit U8977 |
Output Voltage |
Rate: 10mV/A *The output of this device is AC+DC voltage output via the sensor unit. |
Size |
70Wx100Hx53D mm |
Mass (g) |
690 |
Monotaro ID |
52125808 |
Country of Origin |
Japan |
Import Country |
Direct Import from Japan |
Included Components |
Instruction manual x1, mark band x6, precautions for use x1 |
Name of Manufacturer/Packer/Importer |
HIOKI E.E. CORPORATION |
Linearity |
±2 ppm Typical (23℃) |
Cable Length (m) |
10 |
Specifications |
Maximum input current: Within derating range, up to ±420Apeak (design value, below 40℃ and within 20ms) |
Maximum Rated Voltage |
Ground to ground: 1000 V CAT III |
Frequency Characteristic |
Amplitude: DC to 10 MHz, Phase: DC to 1 MHz |
Basic Accuracy |
DC (±0.03% rdg. ±0.002% fs, no phase specification), 45 Hz ≦ f ≦ 66 Hz (±0.03% rdg. ±0.007% fs, ±0.05°), specified up to 1 MHz |
Operating Temperature / Humidity Range |
-40℃ to +85℃, 80% rh or less (no condensation) |
Rated Power (VA) |
Maximum: 6 or less (200A/55Hz measurement, at ±12V power supply) |
Measurable conductor diameter (Φmm) |
24 or less |
Rated Current (A) |
AC/DC 200 |
Offset |
Voltage: ±5 PPM Typical (23℃, no input) |
Looking for bulk pricing or different specifications?